| Customization: | Available |
|---|---|
| After-sales Service: | 1 Year |
| Warranty: | 1 Year |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
| BTS-5V6A Battery Testing System | ||||||||||
| Description | Product Specification | |||||||||
| AC Input | AC 220V±10% 50Hz | |||||||||
| Power | 425 | W | ||||||||
| Resolution | AD: 16bit; DA: 16bit | |||||||||
| Input Impedance | ≥1M | |||||||||
Voltage |
Measuring Range | Charge: 25mV~5V |
Discharge: 0mV~5V |
|||||||
| Discharge Min Voltage | The two up&down head of clamp:1.0V, 1.5m clamp line: 1.5V | |||||||||
| Accuracy | ± 0.05% of FS | |||||||||
| Stability | ± 0.05% of FS | |||||||||
Current |
Each channel output Range | 1st range: 0.5mA~0.1A 2nd range: 0.1A~6A | ||||||||
| Accuracy | ± 0.05% of FS | |||||||||
| CV cut-off current | 1st range:0.02mA 2nd range:12mA | |||||||||
| Stability | ± 0.05% of FS | |||||||||
Power |
Output Power Per Channel | 30W | ||||||||
| Stability | ± 0.1% of FS | |||||||||
Time |
Rise Time | 1ms (0~Full Range) | ||||||||
| Step Time | ≤ (365*24)hour/step Time Format 00:00:00 (h:min:s) | |||||||||
Data Acquisition |
Intervals |
Min. Time interval Δt:100ms | ||||||||
| Min. Voltage interval ΔU: 10mV | ||||||||||
| Min. Current interval ΔI: 1st :0.02mA 2nd:12mA | ||||||||||
| Frequency | 10Hz | |||||||||
Charge |
Mode of Operation | CCC, CVC, CC & CVC,CPC | ||||||||
| End Conditions | Voltage, Current, Test Time, Capacity,-△V | |||||||||
Dischar |
Mode of Operation | CCD, CPD,CRD | ||||||||
| End Conditions | Voltage, Current, Test Time, Capacity,-△V | |||||||||
| Pulse | Charge | CCC,CPC | ||||||||
| Discharge | CCD,CPC | |||||||||
| Min Pulse Width | 500ms | |||||||||
| Pulse no. | Single pulse step support 32 different pulse | |||||||||
| Automated Switch | Automated switch from charge to discharge for each pulse | |||||||||
| End Condition | Voltage, Test Time | |||||||||
| DCIR test | Support custom getting point then DCIR calculate | |||||||||
Cycle |
Cycles | 65535 | ||||||||
| Steps | 254 | |||||||||
| Nested Function | Max three levels of loops | |||||||||
Protective Function |
Software |
Power-off data protection | ||||||||
| Off-line Operating | ||||||||||
| User-defined safety(upper and lower)tolerance of current, voltage and delay time | ||||||||||
| IP Protection level | IP 20 protection level | |||||||||
| Channel Features | Independent pairs of closed loop for constant current source and constant voltage source | |||||||||
| Channels | Independent control | |||||||||
| Detection and Sampling | 4-wire Connecting | |||||||||
| Noise Density | <85dB | |||||||||
| Data Management | MYSQL Database | |||||||||
| Communication Means | TCP/IP Protocol | |||||||||
| SOS | Windows 7 | |||||||||
| Export Formats | EXCEL2003, 2010,TXT | |||||||||
| Communication Interface | Internet | |||||||||
| Number of Channels Per Cabinet | 8 | |||||||||
| Current leakage | 0.0000005mA | |||||||||
| Operating Environment | ||||||||||
| Description | Product Specification | |||||||||
| Operating Temperature | 0ºC~40ºC | |||||||||
| Storage Temperature | -10ºC~50ºC | |||||||||
| Operating Humidity | ≤70% RH | |||||||||
| Storage Humidity | ≤80% RH | |||||||||
| Description | Product Specification | |||||||||
| Holder | Universal Holder | |||||||||
Holder Picture |
||||||||||
| Universal Holder | Alligator Holder | Polymer Holder |
Wiring Terminal | |||||||
| Picture for reference, please confirm the actual product | ||||||||||
| Dimensions (W*D*H)(mm) | 3U(164),480*330*130 | |||||||||
Equipment Picture |
Picture for reference, please confirm the actual product | |||||||||